Multi-Technology Oxygen Sensor | Ntron Microx-OL

Multi-Technology Oxygen Sensor | Ntron Microx-OL

Multi-Technology Oxygen Sensor – Overview

This wall-mounted instrument is designed to measure the oxygen concentration in inert gas supplied under positive pressure. This analyser utilises zirconia or electrochemical technology to give a reliable solution for the measurement of % and ppm oxygen ranges. The zirconia sensor life is typically 3 to 5 years when used within stated specifications. It has fast response times and a long service life. The electrochemical sensor allows for the measurement of background gases containing hydrocarbons. It has virtually no drift, ensuring your measurements are always accurate. It also has integrated audible-visual alarms that you can configure, and is available with multiple measurement ranges to customise your setup.

Multi-Technology Oxygen Sensor  – Features

  • Compact slimline design
  • Integral visual alarm beacon
  • Internal pressure and flow regulation
  • Configurable alarm relays
  • 4-20mA output with RS232 communication
  • LCD + 4 buttons HMI
  • 6mm or 1/4″ compression fittings
  • Zirconia or electrochemical sensor technology options
  • 0 to 1,000ppm, 0 to 25%, 0 to 96% O2
  • Simple to install and easy to operate

Description

Multi-Technology Oxygen Sensor | Ntron Microx-OL

Sensor Zirconia (ZR) Electrochemical (EC)
Measurement Range 0…1,000 ppmV 0…25 %, 0…96 % 0…1,000 ppmV 0…25 %
Accuracy Please see Accuracy Table below for specific point details.
Output Resolution (4…20 mA) 1 ppmV / 0.01 % 0.5 ppmV / 0.01 %
Lower Detection Limit (LDL) 1 ppmV (ppm ranges) / 0.01 % (% ranges)
Sample Flow Rate (application dependent) 100…500 ml/min (250 ml/min optimal) in a vented atmosphere. Set at Factory.
Pressure Range 900…1100 mBarabs (13…16 psi)
Response Time (T90) < 15 seconds @ 25 °C (77 °F) within selected range
Operating Temperature Range -25…+60 °C (-13…+140 °F) 0…+45 °C (+32…+113 °F)
Life Expectancy (application dependent) Up to 5 years Up to 18 months
Humidity 0…95 %rh non-condensing
Shelf Life Unlimited Up to 3 months
Calibration Interval (application dependent) 12 months 3…6 months
Analyzer
Electrical
Display LCD
Output Signal 4…20 mA
Digital Communications RS232
Digital Output Options 1 off relay, dry contact 5 Amp-rated
(2 relay outputs assigned to control light tower indicator)
Electrical Interface 2 x 12-way electrical terminal block
Power Supply 85…230 V AC (50…60 Hz)
Maximum Power Consumption 6 W 2.4 W
Indicators LCD, audible and visual alarms
Mechanical
Sample Connection 6 mm or 1/4” compression fitting
Sample Gas 21 barg (300 psig) maximum
Internal Regulation Sample pressure and flow Factory-set at 0.3 barg (5 psig); 0.55 SCFH (250 ml/m)
Ingress Protection IP65
Housing Material ABS (Rittal mild steel enclosure option available)
Mounting Wall / vertical surface
Compliance
EMC: EN 50270, UKCA

Accuracy Table

Accuracy at standard temperature and pressure (STP)
Range ZR EC
1000 ppm ± 3ppm @ 100 ppm ± 3ppm @ 100 ppm
± 1 ppm @ 10 ppm ± 1 ppm @ 10 ppm
25 % ± 0.03 % @ 1 % ± 0.03 % @ 1 %
± 0.02 % @ 0.1 % ± 0.02 % @ 0.1 %
96% ± 0.5 % @ 20.9 %
± 0.3 % @ 95 %

Multi-Technology Oxygen Sensor | Ntron Microx-OL

Brand

Ntron

Ntron strives to exceed customer expectations by continually innovating better and more efficient gas analysis and control solutions to help maximise process efficiency, improve product quality, protect the health and safety of personnel and the environment, as well as preserve capital equipment and investments.

Reviews

There are no reviews yet.

Be the first to review “Multi-Technology Oxygen Sensor | Ntron Microx-OL”