Description
MMR 620 Micro Ohmmeter
- Measurement of resistive object type:
– welded and soldered connections, equipotential bonding, earth wire
– terminals and connectors, rail welded joints, cables and wires
– measurement 4-pole method
- Measurement of inductive object type:
– coils (motors and transformers), low resistance coils
- Range selection auto-ranging or manual (measurement of inductive object type)
- Selectable measurement mode adjusted to object type:
– for resistive objects – fast mode (3 seconds)
– for inductive objects – long mode with automatic discharge after measurements (or with lower accuracy – shorter mode)
- Selectable measurement mode adjusted to application:
– normal mode – after pushing “START” button
– automatic mode – since test leads are connected to the object meters automatically starts measurement with dual direction current flow and gives average result – eliminates eventual DC voltage on tested object
– continuous mode – every 3 seconds for resistive objects or continuous measurement for inductive objects
- Window mode:
– this option enables setting an upper and a lower limit between which the average measurement result is bound to appear. The results outside of this range will be indicated by two long audible signals and the symbols
- MMR-630 is designated not to be affected by interference up to 20 mV and to show noises from 20 to 100 mV (so in edge conditions even 5 times higher than measured drop of potential)
- Memory store for up to 990 results and a communications facility for transferring data from the device to a PC
Electric security:
– type of insulation: double, according to EN 61010 – 1 and IEC 61557
– measurement category: CAT III 300 V acc. to EN 61010 – 1
– protection class acc. to EN 60529: IP54
Other technical data:
– meter’s power supply: battery package SONEL/Ni-MH 4,8 V
– battery charging time: approx. 2,5 hours
– number of measurements with the current of 10 A: 300
– auto-off time: 120 seconds
– immunity to interference: additional error ≤1% for voltage 50 Hz ≤100 mV RMS
– maximum leads resistance for the 10 A current: 0,1 Ω
– maximum inductance of the tested object: 40H
– accuracy of the test current: ±10%
– resistance measurement time:
– with the selected resistive object type with the bidirectional current: 3 seconds
– with the selected inductive object type, dependent on the resistance
and inductance of the object: a few minutes (max. 10)
– dimensions: 295 x 222 x 95 mm
– weight: approx. 1,7 kg
Rated operational conditions:
– operation temperature: 0…+40°C
– storage temperature: -20…+60°C
MMR-620 microohmmeter conforms the following directives
or standards according to CE requirements.
Directive:
Low Voltage Directive (LVD) 73/23/EEC, 93/68/EEC.
Standards:
EN 61010 – 1:2001 Safety requirements for electrical equipment for measurement, control and laboratory use. General requirements .
EN 61010 – 031:2002: Safety requirements for electrical equipment for measurement, control and laboratory use. Safety requirements for hand-held probe assemblies for electrical measurement and test.
Directive:
Electromagnetic compatibility (EMC) 89/336/EEC, 92/31/EEC, 93/68/EEC.
Standards:
EN 61326:1997+A1:1998+A2:2001 Electrical equipment for measurement, control and laboratory use – EMC requirements.
Resistance measurement
| Range |
Resolution |
| 0…999μΩ |
1μΩ |
10 A |
±(0,25% m.v. + 2 digits) |
| 1,000…1,999 mΩ |
0,001 mΩ |
| 2,00…19,99 mΩ |
0,01 mΩ |
| 20,0…199,9 mΩ |
0,1 mΩ |
1 A |
| 200…999 mΩ |
1 mΩ |
0,1 A |
| 1,000…1,999 Ω |
0,001 Ω |
| 2,00…19,99 Ω |
0,01 Ω |
10 mA |
| 20,0…199,9 Ω |
0,1 Ω |
0,1 mA |
| 200…1999 Ω |
1 Ω |
0,1 mA |
– input impedance of the voltmeter: ≥200 kΩ
“m.v.” – measured value.
Calibration certificates are issued during the production process. New calibration certificates are available upon request at an additional cost.
MMR 620 Micro Ohmmeter
Brand
Sonel
Sonel Details.
Reviews