High-Concentration CO2 Transmitter | GS45

High-Concentration CO2 Transmitter | GS45

R16,207.00 ex vat

High-Concentration CO2 Transmitter – Overview

The High-Concentration CO2 Transmitter is a non-dispersive infra-red (NDIR) CO2 sensor module designed specifically for incubator applications.  The GS45 comes with internal and external mounting options and with pigtail wire connections. The sensor module is powered with DC 12 V and converts the measurement into a linear voltage signal (0 … 5 V corresponding to 0 … 20 vol.% CO2) or RS-485 (Modbus communication corresponding to 0 … 30 vol.% CO2 ). The GS45 also comes with simple maintenance features. The BC wiring connection allows the user to perform a quick background calibration (BC).

High-Concentration CO2 Transmitter – Features

  • non-dispersive infra-red (NDIR) CO2 sensor module
  • pigtail wire connections
  • internal and external mounting options
  • BC wiring allows the user to perform a quick background calibration

Datasheet

Description

High-Concentration CO2 Transmitter | GS45

Specification
Sensing method Non-dispersive infrared(NDIR)waveguide technology
Sampling method Diffusion
Response time(T1/e) 30 sec diffusion time
Measurement range 0 … 20%vol.(0 … 5 V)/ 0 … 30%vol.(RS-485)
Digital resolution 0.01%vol
Repeatability ±0.1%vol.CO2 ±2% of measured value
Accuracy 1, 5 ±0.2%vol.CO2 ±3% of measured value
Pressure dependence 1.6% reading per kPa deviation from normal pressure, 100 kPa
Calibration support BC input to trigger background calibration at 400 PPM(0.04%vol)CO2
(Short circuit to G0 for at least 10 sec)(0 … 5V)/ Calibration by Modbus communication(RS-485)
Voltage output 0 … 5 V corresponding to 0 … 20 vol.% CO2
Output voltage resolution 5 mV or 0.03% CO2
Output load ≥10 kΩ
Communication interface RS-485 Modbus communication
Storage temperature range -40 … +70°C(Non-condensing)
Sensor life expectancy >15 years 2
Maintenance interval Every 12 months
Self-diagnostics Complete function check of the sensor module
Warm-up time ≤1 min
Conformance with the standards Emission:EN61000-6-3:2001;Immunity:EN61000-6-2:2001;RoHS directive 2002 / 95 / EG
Conformance with standards Mechanical shock test IEC 60068-2-27 Test Ea; Random vibration test IEC 60068-2-64 Test Fh
Vibration immunity test severity IEC TR 60721-4-5 table 5:IEC 60721-3-5, Class 5M3(3,6 gRMS)
“Mechanical conditions in road vehicles in areas without well-developed road systems, light-weighted
vehicles, tracked vehicles and self-propelled machines, including installations in places which may be
directly hit by flying stones”
Random vibration test severity IED TR 60721-4-2 table 7:IEC 60721-3-2, Class 2M3(3,2 gRMS)
“Transportation in lorries, trailers and all other kinds of transportation in areas without well-developed
road systems, by trains with shock-reducing buffers and by ships”
IEC 60721-3-2, Class A(1,0 gRMS) “Instrumentation and automation equipment on ships”
Operating temperature range -10 … 50°C
Operating humidity range 0 … 95% RH(Non-condensing)
Operating environment Incubators, residential, commercial and industrial spaces.4
Power input DC 8 … 18 V max rating, stabilized to within 10%(onboard protection circuits)3
Current consumption 40 mA average
<150 mA peak current(Averaged during IR lamp ON, 120 msec)
<250 mA peak power(During IR lamp start-up, the first 50 msec)
Electrical connections4 4xAWG24 cable for power input(Red, Black)and voltage outputs(Blue)and
background calibration(Yellow)
Dimensions 11x7x2.5 cm(Length)x(Width)x(Approximate Height);External dimensions

Note 1: In normal IAQ applications. Accuracy is defined after minimum 3 weeks of continuous operation. However, some industrial applications do require maintenance.
Please, contact us for further information.
Note 2: SO2-enriched environments are excluded.
Note 3: Notice that absolute maximum rating is 14 V, so that sensor can be used with 12 V±10% supply.
Note 4: Different options exist and can be customized depending on the application. Please, find possible options in this document and contact us for further information.
Note 5: Accuracy is specified over operating temperature range. Specification is referenced to certified calibration mixtures.
Uncertainty of calibration gas mixtures(±2% currently)is to be added to the specified accuracy for absolute measurements.

High-Concentration CO2 Transmitter | GS45

Additional information

Options:

0-5V, RS-485

Brand

eYc

Reviews

There are no reviews yet.