Average Flow Thermal Mass Transmitter | FDM06-P

Average Flow Thermal Mass Transmitter | FDM06-P

Average Flow Thermal Mass Transmitter – Overview

This Thermal Mass Transmitter stabilises airflow direction and velocity distribution, leading to improved measurement accuracy and consistent sensor output. It uses a multi-point averaging Pitot tube, which samples airflow across multiple points in the duct to determine a representative average velocity. Suitable for Compressed Air System Management, Optimisation of Air Compressors & Pneumatic Equipment Efficiency,  Air Dryer Flow Control,  Process Gas Consumption Monitoring (N2, Ar, CO2, etc.),  Pipeline Leak Detection & Alerts,  HVAC Duct Monitoring,  Smart Manufacturing Energy Management,  Cleanroom Airflow Monitoring,  Biotech & Pharmaceutical Gas Supply Stability Monitoring,  Food Processing Gas Filling & Packaging Monitoring.

Average Flow Thermal Mass Transmitter – Features

  • Sensor type: Hot-wire sensor
  • High Sensitivity
  • Low Pressure Loss
  • Detects Minute Airflow Changes
  • No Compensation Needed, Direct Flow Output
  • Sizes 2″~12″, Easy Installation
  • ±1.5% F.S. Accuracy, 16 bar Pressure Resistance
  • Analogue, Relay, RS-485 Outputs
  • Suitable for various types of gases (N2, Ar, CO2, etc.)
  • Turndown ratio: 100: 1
  • Output : 4 … 20 mA / 0 … 10 V / Relay / RS-485
  • Accuracy : 0.5 … 60 m/s:±(1.5% of mv + 0.8 m/s)
  • IP rating: IP65

Description

Average Flow Thermal Mass Transmitter | FDM06-P

Input
Sensor type
Hot-wire sensor
Turndown ratio 100: 1
Measuring range 0 … 60 m/s
Output
Output signal
4 … 20 mA / 0 … 10 V / Relay / RS-485
Signal connection 3-wire
Warm-up time 60 sec
Response time t90≦6 sec
Load resistance
Current output: ≦500 Ω
Voltage output: ≧10 KΩ
Communication
Communication methods & protocol
RS-485 Modbus RTU
RS-485 baud rate
9600、19200、38400、57600、115200 bps
Accuracy
Accuracy
0.5 … 60 m/s : ±(1.5% of mv + 0.8 m/s)
Temp. influence 0.2% / °C
Uncertainty of factory calibration ±1%
*The measurement range is defined at the standard condition(1013 mbar, 20°C).
*mv = measured value
Environmental
Medium
Non-corrosion gas
Operating Temp. & Humid.
0 … 50°C / 20 … 90%RH(Non-condensing)
Storage Temp. -20 … +60°C
Operating pressure 16 bar
Electrical
Power supply
DC 24 V ±10%
Current consumption
24 V: 110 mA
Relay capacity
Max current: 6 A
Max voltage: DC 24 V (DC 36 V Max)
Electrical connection
M12 8P connector
Installation
Installation
PT 3/4″ movable thread, PT 1/2″ movable thread
Display
Display readout
0 … 99999999 (Cumulative flow: 8-digit)
0 … 99999 (Instantaneous flow: 5-digit)
Decimal point Button
Sampling time 1 cycle/sec
Flow unit
m/s, ft/s, L/min, m³/min, m³/h, mL, L
m³, ft³, inch³, gal, uk gal
Response time adjustment range
0.5 … 300 sec
Certification
Certification CE
Protection
IP rating
IP65(Housing)
Electrical protection
■ Reverse polarity ■ Over-voltage
Material
Housing
Aluminium alloy / Plastic
Probe SUS316
Note: Design and specifications are subject to change without notice.
Note: If the product has RS-485 function, to avoid damaging product, please make sure the product and the device which connect with RS-485 are on common ground.

Average Flow Thermal Mass Transmitter | FDM06-P

Brand

eYc

Reviews

There are no reviews yet.

Be the first to review “Average Flow Thermal Mass Transmitter | FDM06-P”